Conference Publication Details
Mandatory Fields
Wang, YX;Markham, C;Deegan, C
2019 30TH IRISH SIGNALS AND SYSTEMS CONFERENCE (ISSC)
Assessing the time synchronisation of EEG systems
2019
January
Published
1
1 ()
Optional Fields
SOFTWARE BRAIN P300
This study compared the synchronisation of a medical grade Electroencephalography (EEG) system, the g.Tec, and a consumer grade EEG system, the Emotiv. Data was collected from both systems using the lab streaming layer (LSL). Both EEG systems recorded an electric signal from the surface of a customised gel phantom. The electric signal was generated using a solar cell which was illuminated by a monitor presenting a sequence of black and white images. Test results show that the g.Tec had a mean delay of 51.22 ms from the stimulus onset and the Emotiv had a mean delay of 162.69 ms from the stimulus onset. The result should be taken into account with future ERP studies which will use either the EEG system and the lab streaming layer. The design of this experiment provides a smart way to evaluate the temporal accuracy of other EEG systems.
Grant Details